Abstract

Intravascular ultrasound (IVUS) is an imaging modality more sensitive for defining suboptimal stent deployment issues compared with conventional angiographic imaging. We report on a patient with unstable angina who underwent stent deployment using IVUS guidance. Despite an acceptable angiographic result, intrastent thrombus, incomplete expansion and an edge/marginal dissection were seen by IVUS examination. Both the incomplete expansion and the intrastent thrombus were successfully treated. The dissection, however, was not treated in the catheterization laboratory. This mechanical injury pattern at the stent edge led to vessel occlusion 4 h post-procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.