Abstract

Malfunction of a prosthetic aortic valve is associated with a spectrum of potentially life threatening complications. In addition, the risk of aortic dissection increases following aortic valve replacement, which relates principally to aortic root pathology rather than prosthetic type or its functional status (Circulation 100 (1999) II-287). We report a case in which a high velocity turbulent jet in the proximal aorta, resulting from prosthetic leaflet entrapment caused intimal injury with subsequent type I dissection. This highlights the desirability for regular surveillance following aortic valve replacement, particularly in those patients with an 'at-risk' aorta.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.