Abstract

Efforts have been made to design electrodes that significantly reduce not only the acute and chronic stimulation thresholds, but also attenuate the early peaking phenomenon and polarization. At two voltage levels (2.7 V and 5.4 V, respectively), we evaluated the right ventricular stimulation thresholds obtained with a new, iridium oxide-coated electrode in ten patients who received a VVI pacemaker. Measurements were made at implant and at multiple intervals for 1 year. Pulse width stimulation thresholds at implant were as follow: 0.04 +/- 0.008 msec at 2.7 V, 0.03 +/- 0.004 msec at 5.4 V; values at 2 weeks were 0.14 +/- 0.06 msec at 2.7 V, 0.07 +/- 0.025 msec at 5.4 V; values at 3 months were 0.09 +/- 0.03 msec at 2.7 V, 0.05 +/- 0.01 msec at 5.4 V; values at 1 year were 0.08 +/- 0.02 msec at 2.7 V, 0.04 +/- 0.01 msec at 5.4 V. The maximal increase of 0.11 +/- 0.05 msec occurred at 2.7 V, 2 weeks after implant. Our results indicate that this new electrode provides low acute and long-term stimulation thresholds, as well as an attenuated early peaking phenomenon, being able to stimulate safely at 2.7 V even early after implant.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.