Abstract

152Eu has been standardized by three independent 4π β-γ coincidence counting systems with beta detectors as proportional counter, plastic scintillator and liquid scintillator along with the CIEMAT/NIST method. The average activity concentration by primary methods was linked to key comparison reference value (KCRV) by comparing it with that of 4π γ ionization chamber (GIC) whose calibration factor was determined from the KCRV (BIPM.RI(II)–K1.Eu-152 and CCRI(II)–K2.Eu-152) and deviates from GIC by ± 0.16% indicating good agreement within standard uncertainties.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.