Abstract

201Tl is used in nuclear medicine in cardiac imaging for evaluating the injury level in cardiac muscle at rest and exercise. In this work the activity concentration of a 201 Tl radioactive solution has been absolutely determined using the 4πβ-γ coincidence and sum-peak coincidence methods. The presence of 202Tl radioactive impurity that imposes some difficult in the activity measurements was taken into account in the measurements. In the sum-peak method a planar germanium detector was used. The half-lives were evaluated by the reference source method and the results obtained were (3.033 ± 0.004) d and (12.320 ± 0.163) d, respectively, for 201Tl and 202Tl.