Abstract
A cryogen-free cryostat cooled by a closed cycle cryocooler is compact, can provide uninterrupted long-term operation (up to ten thousand hours) and is suited to temperatures from 3 K to 300 K. Its intrinsic temperature oscillation, however, limits its application in experiments requiring high thermal stability at low temperature (below 77 K). Passive suppression methods are effective but all suffer from drawbacks. We describe a novel, active suppression scheme more efficient than traditional proportional-integral (PI) control. The experimental results show that it can reduce the standard deviation of the temperature oscillation by a further 30% compared with PI feedback. To the best of our knowledge, this is the first time such active suppression of temperature oscillations has been implemented with the cryogen-free cryostat. The results also show, however, that an unwanted lower frequency thermal noise will be generated, which appears to be the limit of the method. Nevertheless, the approach could be used to improve the temperature stability in all cryogen-free cryostats.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.