Abstract
The most widely used architecture in large area amorphous silicon (a-Si) flat panel imagers is a passive pixel sensor (PPS), which consists of a detector and a readout switch. While the PPS has the advantage of being compact and amenable towards high-resolution imaging, reading the low PPS output signal requires external circuitry such as column charge amplifiers that produce additional noise and reduce the minimum readable sensor input signal. This work presents voltage and current mediated active pixel sensor (APS) on-pixel readout circuits for diagnostic medical imaging to minimize the effect of external readout noise sources. Preliminary results indicate excellent APS linearity along with a high programmable gain.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.