Abstract

The elements Fe, Zn, Se, Rb and Co were determined in NBS bovine liver (SRM 1577) by non-destructive activation analysis; ruthenium was used as triple comparator element. The results were compared with the values of the analysis certificate and with those of an analysis based on a multi-element serum standard; the agreement was excellent. The conclusions of the error theory of the comparator method were experimentally confirmed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.