Abstract

A method has been developed to measure the absolute hypersonic absorption of liquids in the frequency range from 0.3 to 3 GHz with 1% accuracy using a pulse transmission technique. Excitation of surface vibrations on small rod-shaped piezoelectric LiNbO3 crystals at variable transducer spacing is combined with a direct parallel RF substitution technique in which a below-cut-off piston attenuator is used as reference. Sensitive superheterodyne detection followed by signal averaging and multiple data recording allow for a low input power level so that heating of the small sample volume (<1 ml) is avoided. The measurements are automatically driven by a process control control.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.