Abstract

We describe the development of a novel instrument intended for the measurement of the acoustical reflection coefficient of materials. The instrument effectively implements a one-port vector-corrected network analyzer in the acoustic, rather than the electromagnetic, domain. Employing the well-documented methods of error correction familiar to microwave engineers, this instrument permits automated measurement of an acoustic impedance presented to a waveguide port. A dual-directional coupler allows a working frequency range of well over an octave. In principle, a set of six couplers would allow measurement from 100 to 50000 Hz.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.