Abstract
Pressure transients occurring inside the eyeball as a consequence of excimer laser ablation of the cornea were measured at various distances along the optical axis, on enucleated porcine eyes. Positive (compressive) pressure spikes up to 90 bars were observed, lasting 100 ns, and developing, as they propagated in the eyeball, satellite rarefaction pulses with negative pressure as high as -40 bar. Such rarefaction wavefronts can trigger the formation of cavitation bubbles, in both the anterior and posterior chamber, depending on the ablation geometry. The potential risks associated with these photoacoustic phenomena, particularly for retina and corneal endothelium, are outlined.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.