Abstract

We measured acoustic output, expressed as the thermal index (TI) and mechanical index (MI), during fetal echocardiography at the time of the first trimester scan. TI and MI were retrieved from the saved displays during gray-mode, high-definition color flow Doppler and pulsed-wave Doppler (tricuspid flow) ultrasound examinations of the fetal heart and from the ductus venosus assessment. A total of 399 fetal cardiac examinations were evaluated. There was a significant increase in TI values from B-mode studies (0.07 ± 0.04 [mean ± SD]) to color flow mapping (0.2 ± 0.0) and pulsed-wave Doppler studies (0.36 ± 0.05). The TI from ductus venosus assessment (0.1 ± 0.01) was significantly lower than those from Doppler examinations of the heart. MI values from B-mode scans (0.65 ± 0.12) and color flow mapping (0.71 ± 0.11) were comparable, although different, and both values were higher than those from pulsed-wave Doppler tricuspid evaluation (0.39 ± 0.03). There were no differences in MI values from power Doppler assessment between the tricuspid flow and ductus venosus. Safety indices were remarkably stable and were largely constant, especially for color Doppler (TI), tricuspid flow (MI) and ductus venosus assessment (TI, MI). We acquired satisfactory Doppler images and/or signals at acoustic levels that were lower than the actual recommendations and never reached a TI of 0.5.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.