Abstract

An acoustic isolation vessel has been developed to measure the background noise in microphones. The test microphone is installed in an inner vessel, which is suspended within an outer vessel, and the intervening air space is evacuated to a high vacuum. An analytical expression for the transmission coefficient is derived, based on a five-media model, and compared to experiment. At an isolation vacuum of 5×10−6 Torr the experimental transmission coefficient was found to be lower than −155 dB at frequencies ranging from 40 to 1200 Hz. Measurements of the A-weighted noise levels of commercial condenser microphones of four different sizes show good agreement with published values.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.