Abstract
Evidence is presented of acoustic cavitation generated by a Dornier extracorporeal shockwave lithotripter. Using x-ray film, thin aluminium sheets, and relatively thick metal plates as targets, evidence of liquid jet impacts associated with cavitation bubble collapse was observed. The jet impact was violent enough to puncture thin foils and deform metal plates. Furthermore, numerous jet impacts were generated over a volume of greater than 200 cm 3. It is likely that such violent cavitation will also occur in tissue, and observed biological effects (e.g. renal calculus disintegration and tissue trauma) may be related to cavitation damage.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.