Abstract

A survey was conducted of acoustic noise levels in magnetic resonance (MR) imaging systems. Static magnetic flux and radio-frequency pulses did not affect the function of shielded sound-pressure detection equipment. Noise levels were measured at bore isocenter during a variety of imaging sequences in six MR imaging systems with magnetic fields of 0.35-1.5 T. Measured noise ranged from 82 to 93 dB on the A-weighted scale and from 84 to 103 dB on the linear scale. Noise levels increased during sequences employing thinner section thickness and shorter repetition and echo times and were independent of field strength. Gradient-coil noise in MR imaging is an annoyance but is well within safety guidelines.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.