Abstract

Glow discharge conditioning (GDC) has long been accepted as one of the basic wall conditioning techniques for achieving ultrahigh vacuum in an unbaked chamber. As a part of this fundamental experimental study, a test chamber has been fabricated from stainless steel 304 L with its inner surface electropolished on which a detailed investigation has been carried out. Both helium and hydrogen gases have been employed as discharge cleaning medium. The discharge cleaning was carried out at 0.1 A /m2 current density with working pressure maintained at 1.0 × 10 -2 mbar. It was experimentally observed that the pump-down time to attain the base pressure ~10 -8 mbar was reduced by 62% compared to the unbaked chamber being pumped to this ultimate vacuum. The results were similar irrespective of whether the discharge cleaning medium is either hydrogen or helium. It was also experimentally established that a better ultimate vacuum could be achieved as compared to theoretically calculated ultimate vacuum with the help of discharge cleaning.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.