Abstract

We have measured the acetylene concentration in the exhaled breath of 40 volunteers (31 non-smokers, nine smokers) using near-infrared cavity ring-down spectroscopy. The acetylene levels were found to be the same as in ambient air for non-smokers, whereas elevated levels were observed for smokers. Real-time measurements with sub-second time resolution have been applied to measure the elimination kinetics of acetylene in breath after exposure to tobacco smoke. Three exponential time constants can be distinguished from the data and these can be used to define the residence times for different compartments, according to the multi-compartment model of the human body.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.