Abstract

Water pollution due to toxic metals may pose a significant effect on the environment and the biota indicating necessity of careful management of their concentration in the system. On the other hand, a simplified and useable sample preparation technique is required for their trace analysis. In this report, a simple solid phase extraction procedure was developed for fast extraction and preconcentration of Cu(II) and Cr(III) from waters using acetic acid modified zinc oxide nanoparticles prior to flame atomic absorption spectrometry. The extraction procedure was optimized in respect to sample volume, pH, adsorbent dose, eluent, and extraction time. Ionic strength and co-existing ions did not exhibited any significant effect on the extraction. Under optimum conditions, with an enrichment factor of 40 and recoveries of over 95%, the limits of detection for both analytes were in the low µg L−1 range. Intra-day and inter-day precision were at about 3% to 4%. The method was applied for trace determination of the analytes in real water samples.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call