Abstract

Acetabular component deformation secondary to forces encountered during insertion is a potential consequence of the press-fit technique. This study characterized the stiffness of Pinnacle 100 cups (DePuy, Warsaw, Ind) via mechanical testing and used this information with intraoperative measurements of cup deformation to calculate the in vivo forces acting on cups inserted during hip arthroplasty in 21 patients. We found that 90.5% of cups had measurable compression deformity, averaging 0.16 ± 0.16 mm. The corresponding forces acting on these cups averaged 414 ± 421 N. For hard-on-hard bearing surfaces, such in vivo deformation of acetabular shells may result in negative clinical consequences such as equatorial loading with increased wear and potential seizing of components, chipping of ceramic inserts, or locking mechanism damage.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.