Abstract

Accurate determination of technetium-99 (Tc-99) is very important because any overestimation will cause the examined radioactive wastes to be categorized into super C class, which dramatically increases the cost of waste management. Herein, we demonstrated that by adopting the analytical method comprising TEVA resin pretreatment and ICP-MS measurement, the determined Tc-99 concentrations in representative waste stream samples from the Lan-Yu low-level radioactive waste temporary storage site in Taiwan were approximately two orders of magnitude lower than those determined from the beta radiation measurement using a low background liquid scintillation counter. Two important concerns emerged from our results. First, severe interferences from other nuclides residing in the matrix considerably affect the determination of Tc-99, even when a low background liquid scintillation counter was used. Second, the currently used Tc-99/Cs-137 scaling factor should be carefully revised, or it might lead to a considerable overestimation of the Tc-99 concentration.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.