An approach to improve the accuracy of submicron particle size distribution measurement using a cuvette photocentrifuge (Horiba CAPA-700) has been applied to a commercial alumina powder and a standard quartz powder. This approach incorporates a correction for the light scattering from small particles, an improved raw data smoothing routine and a correction for the radial dilution effect. The results for the commercial alumina powder show very good correspondence with the supplier's X-ray sedimentometer data when corrected for light scattering and radial dilution; and allow a more complete data collection in the 0.3–0.03 μm range. For the standard quartz powder (BCR 66) a less satisfactory result is found, with the light-scattering corrected data giving a much finer particle size distribution than expected for the standard powder. Two possible reasons for this discrepancy, the nonsphericity of the quartz powder and an imaginary part of the refractive index of quartz due to impurities, are discussed along with the limitations of the approach for accurate submicron particle size measurement of polydispersed irregularly shaped particles.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.