Abstract

After an explosion of pentaerythritol tetranitrate (PETN), its degradation products pentaerythritol trinitrate (PETriN), dinitrate (PEDiN) and mononitrate (PEMN) were detected using liquid chromatography–atmospheric-pressure chemical-ionization–mass spectrometry (LC–APCI–MS). Discrimination between post-explosion and naturally degraded PETN could be achieved based on the relative amounts of the degradation products. This information can be used as evidence when investigating a possible relationship between a suspect and a post-explosion crime scene. The present work focuses on accurate quantitation of PETN and its degradation products, using PETriN, PEDiN and PEMN standards specifically synthesized for this purpose. With the use of these standards, the ionization behavior of these compounds was studied, and a quantitative method was developed. Quantitation of PETN and trace levels of its degradation products was shown to be possible with accuracy between 85.7% and 103.7% and a precision ranging from 1.3% to 11.5%. The custom-made standards resulted in a more robust and reliable method to discriminate between post-explosion and naturally-degraded PETN.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.