Abstract
To avoid the need for radiological control in the assessment of the proper location of central venous catheters (CVC), a particular use of endocavitary electrocardiography (EC-ECG) was proposed 10 years ago. The aim of this study is to reassess our experience with this method. EC-ECG assumes that the CVC, when filled with normosaline and connected to a standard electrocardiograph, behaves like an exploring electrode. The approach of the catheter tip to the right atrium is then detected by a slightly increasing negative P wave. When the tip reaches the exact level of the sinus node, the P wave suddenly deepens. After a preliminary test of the reliability of the technique versus the standard method in 50 CVC placements verified by both EC-ECG and chest x-ray, we have placed 807 CVCs in children using EC-ECG only. There have been no false-positive and one false-negative test result (lead connector misplacement). In 17 cases in which intrinsic deflection was not detected, the catheter tip was found to be wrongly positioned; all the remainder CVCs have been successfully positioned. For 10 years this technique has proved to be a simple, safe, quick, inexpensive and highly reliable method to assess the correct positioning of the CVC.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.