Abstract

This study seeks to demonstrate the ability of portable X-ray fluorescence (PXRF) to estimate concentrations of K, Ca, and Fe in sediments and archaeological ceramics under controlled conditions. After a discussion of the potential confounding factors in PXRF use, a protocol which attempts to address these issues through repeated measurement, calibration, and re-sampling is detailed. Data generated using this protocol are then tested for accuracy and repeatability. PXRF is argued to be able to produce accurate estimates of K provided the suggested protocol is used, and able to produce repeatable estimates of K and Ca under these same conditions. Other experimental conditions tested failed to produce accurate and repeatable results. Fe results are found to be problematic given the calibration standards used here.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.