Abstract

In a terahertz (THz) time-domain spectroscopic system, the THz beam waist radius can be measured by eclipsing the beam waist with a semi-infinite mask while monitoring signal variation (knife-edge technique). However, the obtained beam waist radius is affected by the spatial distribution of the detector sensitivity. This effect was confirmed by calculating the refractive-index corrections required when inserting a sample into the beam waist with and without consideration of the detector sensitivity distribution and comparing them to the experimental results. The real beam waist radius was 2 times larger than that obtained by simply differentiating the signal variation in a system comprising optics symmetrical about the beam waist. This effect must be considered in obtaining the correct THz beam profile by the knife-edge technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.