A synchronous measurement method that can be applied for the accurate determination of periodic signals is presented. The method relies on adaptation of the measurement time to the period of the measurement signal. It is based on dual-slope integration performed according to the synchronization signal (input signal). The first phase of integration is fully synchronized with the input signal. The time of the first phase is also measured. The output voltage of the integrator depends directly on the input signal and the time of the first integration phase. Measurement values are the results of arithmetic operations between the time of the second phase of integration and the measuring time (first phase). With this method, an accuracy of 0.1% can be achieved.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.