Abstract
An accurate technique is developed to correct the intensity variation in the magnetic resonance images acquired with an endorectal surface coil for prostate imaging. The technique relies on reconstructing the in vivo size, shape, and location of the coil during imaging, wherefrom a pixel-by-pixel correction is applied to the coil's sensitivity profile in accordance with the Biot-Savart law. The method is evaluated by a test-object containing multiple layers of beads embedded in gelatine. Qualitatively, the improvement in visualizing regions of interest after corrections is obvious. Quantitatively, the local SNR shows an improvement by a factor of 1.5-2. Several human studies are performed. The technique shows remarkable improvement in visualizing the prostate region.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.