Abstract

A new method accurately (2 msec, 0.2 beats/minute) measures the cycle length of artificial pacemakers with the conventional electrocardiograph and the 60-Hz AC interference artifact. The measurements are easily performed by a technician in less than 10 minutes and, in the absence of pacemaker clinic facilities, may be used to predict pacemaker battery failure in most currently manufactured models.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.