Abstract

The aim of this study was to propose an improved method for accurate dialysis dose evaluation and extrapolation by means of Kt/ V from online UV-absorbance measurements for real time and continuous treatment monitoring. The study included a total of 24 treatments from ten uremic patients, seven of whom were male and three females. All patients were on chronic thrice-weekly hemodialysis therapy. The study included both stable and unstable treatments. A known signal processing algorithm, Levenberg-Marquardt, and the newly developed SMART were utilized for the removal of disturbances not relevant for dialysis dose evaluation. Finally, the results were compared with the Kt/ V values based on the blood samples. The new data processing algorithm, SMART, removes disturbances, helps estimate the online Kt/ V with significant precision increase and without any time delay, and more effectively predicts the end Kt/ V for the treatment than the known algorithms.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.