Abstract
We demonstrate two robust protocols for determining the equilibrium surface tension (EST) values with area perturbation tests. The EST values should be indirectly determined from the dynamic surface tension (DST) values when surface tension (ST) values are at steady-state and stable against perturbations. The emerging bubble method (EBM) and the spinning bubble method (SBM) were chosen, because, with these methods, it is simple to introduce area perturbations while continuing dynamic tension measurements. Abrupt expansion or compression of an air bubble was used as a source of area perturbation for the EBM. For the SBM, changes in the rotation frequency of the sample solution were used to produce area perturbations. A Triton X-100 aqueous solution of a fixed concentration above its critical micelle concentration (CMC) was used as a model surfactant solution. The determined EST value of the model air/water interface from the EBM was 31.5 ± 0.1 mN·m-1 and that from the SBM was 30.8 ± 0.2 mN·m-1. The two protocols described in the article provide robust criteria for establishing the EST values.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.