Abstract

A combined fire assay-wet chemical procedure for the accurate and rapid determination of platinum, palladium, gold and silver is described. The sample is fused with lead flux and the lead button is parted with perchloric acid after heating at 180–190 °C. After dilution of the parting solution with water, a formic acid reduction step is applied for the selective group precipitation of the four noble metals. After filtration, the precipitate is dissolved in aqua regia and the noble metals are determined by atomic absorption spectrometry. The proposed method was applied to a variety of sample types and, compared with the classical lead cupellation method, showed greater accuracy and precision in all instances.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.