Abstract

A precise and accurate potentiometric titration of 2–25 mg of uranium is described. The uranium is reduced to U(IV) according to the method of Eberle et al. [3], and titrated with 0.05 N potassium dichromate, using a platinum indicator electrode. During the sample preparation the walls of the titration vessel are cleaned by centrifugation. To avoid overshoot of the set point a special differentiator is described, that interrupts the titration until equilibrium is reached. The precision of the method is 0.02%, the accuracy is better than 0.04% rel. Each titration takes 5 min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.