We have shown that the Hewlett-Packard 47201A ear oximeter measures arterial O2 saturation within 95 per cent confidence limits of +/- 4 per cent when arterial blood saturation is more than 65 per cent, but at lower saturations the oximeter consistently provides a reading that is too low. The oximeter is sensitive to carboxyhemoglobin, progressively overestimating arterial saturation as carboxyhemoglobin concentration increases from 0 to 18 per cent. The time response is exponential, with a normal time constant of 3 sec, but this is halved in the fast mode or operation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.