Abstract

The performance of six self-monitoring blood glucose (SMBG) machines (Accutrend ®, Reflolux S ®, Companion 2 ®, Glucometer GX ®, Glucometer IV ® and One Touch II ®) were examined using venous blood samples from 88 patients. Whole blood glucose (BG) values were measured by four machines from each brand. Machine-generated whole blood glucose (BG) values were corrected before comparison with laboratory plasma glucose values, measured by a glucose oxidase method. Based on error grid analysis, most of the corrected machine-generated BG values were clinically acceptable. Accutrend ®, Glucometer IV ® and Companion 2 ® showed the greatest consistency between machines of the same brand. Over 80% of corrected BG values generated by Glucometer IV ® fell within ± 10% of the reference values. One Touch II ® yielded the most reproducible results with a mean CV of 2.7% and was considered the most user friendly machine. More studies are required to examine the performance of these machines in the hands of patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.