Abstract
Background: Weight bearing computed tomography (WBCT) utilizes cone beam CT technology to provide assessments of lower limb joint structures while they are functionally loaded. Grey-scale values indicative of X-ray attenuation that are output from cone beam CT are challenging to calibrate, and their use for bone mineral density (BMD) measurement remains debatable. To determine whether WBCT can be reliably used for cortical and trabecular BMD assessment, we sought to establish the accuracy of BMD measurements at the knee using modern WBCT by comparing them to measurements from conventional CT.Methods: A hydroxyapatite phantom with three inserts of varying densities was used to systematically quantify signal uniformity and BMD accuracy across the acquisition volume. We evaluated BMD in vivo (n = 5, female) using synchronous and asynchronous calibration techniques in WBCT and CT. To account for variation in attenuation along the height (z-axis) of acquisition volumes, we tested a height-dependent calibration approach for both WBCT and CT images.Results: Phantom BMD measurement error in WBCT was as high as 15.3% and consistently larger than CT (up to 5.6%). Phantom BMD measures made under synchronous conditions in WBCT improved measurement accuracy by up to 3% but introduced more variability in measured BMD. We found strong correlations (R = 0.96) as well as wide limits of agreement (-324 mgHA/cm3 to 183 mgHA/cm3) from Bland-Altman analysis between WBCT and CT measures in vivo that were not improved by height-dependent calibration.Conclusion: Whilst BMD accuracy from WBCT was found to be dependent on apparent density, accuracy was independent of the calibration technique (synchronous or asynchronous) and the location of the measurement site within the field of view. Overall, we found strong correlations between BMD measures from WBCT and CT and in vivo measures to be more accurate in trabecular bone regions. Importantly, WBCT can be used to distinguish between anatomically relevant differences in BMD, however future work is necessary to determine the repeatability and sensitivity of BMD measures in WBCT.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.