Abstract
To assess the accuracy of left ventricular volume measured by the intraventricular balloon method, we measured the space between the endocardium and the gradually inflated balloon in six Formalin-fixed canine left ventricles. The space was as small as 1-1.5 ml for intraballoon pressure of 100-200 mmHg and tended to increase to 2 ml with decreases in pressure to 50 mmHg.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: American Journal of Physiology-Heart and Circulatory Physiology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.