Abstract
The primary objective of the present study was to evaluate the accuracy of a new robotic device when coupled with intraoperative flat-panel CT guidance. Screws (D8-S1) were implanted during two separate cadaver sessions by coupling the Rosa(®) Spine robot with the flat-panel CT device. Of 38 implanted screws, 37 (97.4 %) were fully contained within the pedicle. One screw breached the lateral cortical of one pedicle by <1 mm. The mean ± SD accuracy (relative to pre-operative planning) was 2.05 ± 1.2 mm for the screw head, 1.65 ± 1.11 for the middle of the pedicle and 1.57 ± 1.01 for the screw tip. When coupled with intraoperative flat-panel CT guidance, the Rosa(®) Spine robot appears to be accurate in placing pedicle screws within both pedicles and the vertebral body. Large clinical studies are mandatory to confirm this preliminary cadaveric report.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.