Abstract

The performance of the VITEK® 2 system for direct rapid identification and antimicrobial susceptibility testing of the bacteria responsible for blood infections was determined. The isolates studied included 166 Gram-negative rods and 74 Gram-positive cocci from inpatients. Specially treated monomicrobial samples from positive blood culture bottles were directly inoculated into the VITEK 2 system and the results were compared with those from cards inoculated with standardized bacterial suspensions. Compared with the standard method, 95.8% of Gram-negative rods were correctly identified by VITEK 2 and the overall level of agreement between the two methods in susceptibility testing was 92.0%. For Gram-positive bacteria, 89.2% were correctly identified by VITEK 2 and susceptibility testing revealed an overall agreement rate of 91.3%. These results suggest that VITEK 2 cards inoculated with fluids sampled directly from positive blood culture bottles are suitable for speedy identification and susceptibility testing of Gram-negative bacilli and Gram-positive cocci.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call