Abstract

Introduction: Ultrasound facilitates neuraxial puncture in obese parturients. Unfortunately, the widespread adoption of neuraxial ultrasound may be limited by the lack of technical expertise or the limited availability of the equipment. A wireless portable ultrasound device (Accuro, Rivanna Medical, Charlottesville, VA) with automated pattern recognition software (SpineNav3DTM technology) has been introduced to obtain the automated real-time identification of interspaces and epidural depth. The primary objective of this study was to assess the accuracy of the SpineNav3DTM ultrasound technology (Accuro) in estimating the epidural space depth compared to the standard ultrasound examination in pregnant obese patients. The secondary aim was to compare the ultrasound measurements with the measured needle depth during epidural and spinal insertion. Methods: The study was conducted at S Gerardo Hospital, Monza, Italy from March 2021 to April 2021. Obese laboring women requesting epidural analgesia or undergoing elective cesarean delivery under spinal anesthesia were recruited. All the subjects had their lumbar area scanned for the measurements of the depth of the epidural space by the SpineNav3DTM ultrasound technology and by the standard US and then both compared with the needle insertion depth in a double-blind fashion. Results: Forty-eight women were enrolled in the study. There was agreement (±0.25 cm) between the epidural depth (in cm) measured with the Accuro, versus the standard ultrasound. There was a significant difference between the mean depth of epidural space s measured by Accuro or Standard US and needle insertion depth (P Conclusions: The handheld ultrasound system with 3D spine navigation technology can automatically identify and measure the epidural depth with the same accuracy as the standard ultrasounds in obese pregnant women.

Highlights

  • Ultrasound facilitates neuraxial puncture in obese parturients

  • The primary objective of this study was to assess the accuracy of the SpineNav3DTM ultrasound technology (Accuro) in estimating the epidural space depth compared to the standard ultrasound examination in pregnant obese patients

  • All the subjects had their lumbar area scanned for the measurements of the depth of the epidural space by the SpineNav3DTM ultrasound technology and by the standard US and both compared with the needle insertion depth in a double-blind fashion

Read more

Summary

Introduction

Ultrasound facilitates neuraxial puncture in obese parturients. the widespread adoption of neuraxial ultrasound may be limited by the lack of technical expertise or the limited availability of the equipment. A wireless portable ultrasound device (Accuro, Rivanna Medical, Charlottesville, VA) with automated pattern recognition software (SpineNav3DTM technology) programmed to calculate the depth to the epidural space and identify bony landmarks has been introduced in order to require less operator interpretation of ultrasound images while offering automated real-time identification of interspaces and epidural depth This device has been reported to accurately determine the depth to the epidural space and to identify the correct interspace in normal-weight women receiving labor epidural analgesia [8] and to be capable of measuring the epidural depth with the same accuracy as the standard ultrasounds in normal weight term pregnant volunteers [9]

Objectives
Methods
Results
Conclusion