Abstract

Rapid functional assays have been proposed to overcome the limitations of washed platelet assays in the work-up of patients with suspected heparin-induced thrombocytopenia (HIT). Data on the diagnostic accuracy are, however, scarce and conflicting. We aimed to study the diagnostic accuracy of a rapid, flow cytometer-based assay and to explore sources of variability. Frozen serum samples of 103 consecutive patients, evaluated for suspected HIT at our institution in 2017, and characterized with 4Ts score, IgG-PF4/heparin ELISA (GTI), HemosIL®Acustar (IgG), as well as heparin-induced platelet activation test (HIPA), were further tested using HIT Confirm, determining P-selectin release of donor platelets after incubation with patient's serum. The diagnosis of HIT was defined as a positive HIPA result. HIT was confirmed in 15 out of 103 patients corresponding to a prevalence of 14.6%. HIT Confirm was positive in 11 patients (10.7%), negative in 88 patients (85.4%), and inconclusive in 4 patients (3.9%). According to the intention-to-diagnose principle, the number of true positives was 9, the number of true negatives 83, the number of false negatives was 6, the number of false positives 5. This corresponds to a sensitivity of 60.0%, and a specificity of 94.3%. Modifications of the test did not improve sensitivity. The rapid, flow cytometer-based assay HIT Confirm is able to verify HIT in positive patient samples but cannot rule-out HIT in clinical practice. Other rapid functional assays shall be studies in appropriately designed diagnostic accuracy studies.

Highlights

  • Because of the complex mechanism of heparin-induced thrombocy­ topenia (HIT), the diagnostic work-up is challenging

  • We hypothesized that various test modifications might improve the diagnostic performance of HIT Confirm: (1) use of washed platelets rather than platelet-rich plasma (PRP), (2) using platelets of a second donor, and (3) filtrating the patient’s serum

  • HIT was confirmed by a positive heparin-induced platelet activation test (HIPA) test in 15 patients (14.6%), a number that was higher than expected

Read more

Summary

Introduction

Because of the complex mechanism of heparin-induced thrombocy­ topenia (HIT), the diagnostic work-up is challenging. HIT is an immu­ nological disorder caused by the development of antibodies that recognize platelets factor 4 (PF4) complexed with polyanions such as heparin (anti-PF4/ H abs). These immune complexes bind to FcγIIAreceptors, trigger activation of platelets, support derivation of platelet microparticles, and, subsequently, accelerate thrombin generation [1]. Well-known clinical consequences are thrombocytopenia and massive thromboembolism. Immunoassays such as enzyme-linked immunosor­ bent assay (ELISA) or chemiluminescence immunoassay (CLIA) [2] can quantify anti-PF4/H antibodies with high sensitivity [3]. Functional assays are needed to demonstrate anti-PF4/H anti­ bodies’ capability to activate platelets [1]

Objectives
Methods
Results
Discussion
Conclusion