Abstract

The accuracy of estimating the duration of single-cycle laser pulses with centre wavelengths of 910 and 780 nm is analysed using a single-shot second-order intensity autocorrelator. It is shown that estimates of the single-cycle pulse duration with an error of less than 5 % require the use of KDP crystals with a thickness of no more than 10 μm for the second harmonic generation. At the same time, to estimate the duration of a transform-limited pulse containing ten optical cycles, the crystal thickness can be up to 1 mm. In this case, at an optimum angle of convergence of the first-harmonic beams, the duration estimation error is less than 2 %.