Abstract

Blood and tissue velocity are measured and analysed in cardiac, vascular and other applications of diagnostic ultrasound (US). An error in system calibration is a potential risk for misinterpretation of the measurements. To determine the accuracy in velocity calibration, we tested three common commercial US systems using a Doppler string phantom. We tested pulsed and continuous-wave Doppler modes for velocities relevant to both cardiac blood flow and tissue-velocity estimation. The US systems were tested with settings and transducers commonly used in cardiac applications. One system consistently overestimated velocity by about 5%, whereas the other two systems were quite accurate in velocity estimation. These findings emphasize the importance of continuous quality control of US equipment. (E-mail: andrew.walker@ltvastmanland.se)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.