Abstract
The accuracy of time spread measurements of fast photomultipliers has been investigated, using single photoelectrons. The effect of the finite light pulse width on the measurement accuracy is determined and discussed. Experimental data have been obtained on a special measuring system for light pulse widths ranging from 200 ps to 10 ns, using fast photomultipliers 8850 and C31024 with optimized operating conditions for minimum transit time spread. A modified exponential function expression and curve-fitting parameters are given, which fit closely the experimentally obtained data over a wide dynamic range of light pulse widths.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.