Abstract
Our goal was to determine whether Doppler ultrasonographic estimation of fetal right ventricular output was accurate enough to evaluate fetal cardiac performance before the technique can be applied with confidence to the human fetus. The accuracy of this method has not been validated experimentally. We compared the right ventricular output measured by Doppler echocardiography with that determined from a transit-time ultrasonic flowmeter placed around the fetal pulmonary trunk and to that determined by microsphere technique. We obtained 72 Doppler flow measurements and 16 microsphere flow measurements in six fetuses. There was a fair correlation between cardiac output estimated by Doppler echocardiography performed transabdominally and that determined by a transit-time flowmeter (r = 0.76). The correlation was excellent when Doppler echocardiography was performed by the trans-uterine approach (r = 0.94). There was also an excellent correlation between right ventricular output determined by a transit-time flowmeter and that by microsphere technique (r = 0.94). Doppler echocardiographic estimation of fetal right ventricular cardiac output is thus feasible and accurate.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.