Abstract

ObjectivesWe sought to explore the accuracy of remote chest X-ray reading using hands-free, wearable technology (Google Glass, Google, Mountain View, California). MethodsWe compared interpretation of twelve chest X-rays with 23 major cardiopulmonary findings by faculty and fellows from cardiology, radiology, and pulmonary-critical care via: (1) viewing the chest X-ray image on the Google Glass screen; (2) viewing a photograph of the chest X-ray taken using Google Glass and interpreted on a mobile device; (3) viewing the original chest X-ray on a desktop computer screen. One point was given for identification of each correct finding and a subjective rating of user experience was recorded. ResultsFifteen physicians (5 faculty and 10 fellows) participated. The average chest X-ray reading score (maximum 23 points) as viewed through the Google Glass, Google Glass photograph on a mobile device, and the original X-ray viewed on a desktop computer was 14.1±2.2, 18.5±1.5 and 21.3±1.7, respectively (p<0.0001 between Google Glass and mobile device, p<0.0001 between Google Glass and desktop computer and p=0.0004 between mobile device and desktop computer). Of 15 physicians, 11 (73.3%) felt confident in detecting findings using the photograph taken by Google Glass as viewed on a mobile device. ConclusionRemote chest X-ray interpretation using hands-free, wearable technology (Google Glass) is less accurate than interpretation using a desktop computer or a mobile device, suggesting that further technical improvements are needed before widespread application of this novel technology.