Abstract

To determine the accuracy of pulse oximetry measurement of heart rate in the neonatal intensive care unit. Stable preterm infants were monitored with a pulse oximeter and an ECG. The displays of both monitors were captured on video. Heart rate data from both monitors, including measures of signal quality, were extracted and analysed using Bland Altman plots. In 30 infants the mean (SD) difference between heart rate measured by pulse oximetry and electrocardiography was -0.4 (12) beats per minute. Accuracy was maintained when the signal quality or perfusion was low. Pulse oximetry may provide a useful measurement of heart rate in the neonatal intensive care unit. Studies of this technique in the delivery room are indicated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.