Abstract

To evaluate the accuracy of noninvasive blood pressure (NIBP) measurement at the dependent- and nondependent arms in the lateral position, using invasive blood pressure (IBP) as reference. This prospective observational study included 42 adult patients undergoing surgery in the lateral position. Paired readings of IBP and NIBP were obtained at either arm. The accuracy of both arms in detecting mean arterial pressure (MAP) <70 mmHg was evaluated using the area under the receiver operating characteristic curve (AUC). The agreement between the IBP and NIBP was evaluated using the Bland-Altman and error grid analyses. We analyzed 350 and 347 paired readings at the dependent- and nondependent arms. The AUC for detecting hypotension was comparable in both arms. The negative and positive predictive values (95% confidence interval) were 100% (99-100%) and 24% (14-34%), respectively for the dependent arm at cutoff value MAP ≤86 mmHg; and were 99% (96-100%) and 21% (13-30%), respectively for the nondependent arm at cutoff value MAP ≤75 mmHg. The mean bias for MAP was -6.0 ± 9.1 and 6.3 ± 10.1 mmHg; and for systolic blood pressure was 0.3 ± 11.6 and 13.2 ± 12.6 mmHg, in the dependent- and nondependent arm, respectively. Error grid analysis showed that the proportions of paired MAP readings in risk zone A were 71 and 82% in the dependent- and the nondependent arms, respectively. In the lateral position, the NIBP readings at both arms are not interchangeable with the corresponding IBP readings. However, NIBP measurement at both arms can be used to accurately rule out hypotension.