Abstract

Purpose To assess the accuracy and variability in pachymetry measurements obtained by Orbscan and by DGH ultrasound in primary laser in situ keratomileusis (LASIK) and LASIK enhancement procedures. Setting Gimbel Eye Centre, Calgary, Alberta, Canada. Methods A retrospective analysis of 906 consecutive primary LASIK and 183 consecutive LASIK enhancement procedures for which preoperative DGH ultrasound and Orbscan pachymetry measurements were obtained. The theoretical residual corneal thickness was compared to measurements by both instruments in 60 eyes that had primary procedures and enhancements. Results In primary LASIK eyes, the DGH ultrasound measurements were thicker than the Orbscan measurements by a mean of 18.4 μm ± 17.4 (SD). The DGH ultrasound measurements were thicker than the Orbscan measurements by a mean of 50.1 ± 40.7 μm in preenhancement pachymetry measurements. The DGH ultrasound measurements were consistent with theoretical residual corneal thickness, 493.0 ± 42.0 μm versus 487.0 ± 31.0 μm ( P = .65), while Orbscan measurements were statistically less than the theoretical residual corneal thickness, 431.0 ± 42.0 μm versus 468.0 ± 39.0 μm ( P = .0001). Conclusion DGH ultrasound was a more accurate measurement of corneal pachymetry than Orbscan. The discrepancies between DGH ultrasound and Orbscan pachymetry measurements were more prominent in eyes that had had LASIK.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.