Abstract

To analyze the accuracy of growth-plate thickness measurements detected on 1.5-T and 7-T MR images using histology sections as a standard of reference. Four defrosted pig tibiae were 1.5-T MR scanned and one fresh tibia was 7-T MR scanned. The height of the growth plate was measured and compared to histology. Histology measurements showed a mean growth plate thickness of 467μm (SD = 82.2). The mean growth plate thickness measured in the 7-T MR images was 465μm (SD = 62.2) and 1,325μm (SD = 183.5) on 1.5-T MR measurements. We found a better correspondence between the growth plate thickness measured on the 7-T MR and histology samples compared to 1.5T. The growth plate can be identified and measured with high accuracy using 7-T MR. 1.5-T MR can only describe some morphological characteristics.