Abstract

The relative light output per Gy in polystyrene for roentgen beams of 6 and 42 MV and electrons between 2.2 and 34.5 MeV relative to 60Co gamma radiation is reported for different kinds of LiF dosemeters. The distribution of the absorbed dose inside a 0.25 and 0.4 mm thick LiF-teflon disc surrounded by polystyrene and irradiated with 60Co, 42 MV roentgen radiation and 39 MeV electrons was measured using 0.01 and 0.02 mm thick Lif-teflon discs. The measurements show that the absorbed dose distribution in the dosemeter depends on the energy of the radiation. When flat dosemeters were used, differences between the signals measured at the two orientations possible during read-out could easily amount to several per cent, and for this reason 0.4 mm and 0.5 mm LiF-Teflon discs were not trusted when the highest accuracy was required. The cavity theory by Burlin does not account for the phenomena caused by differences in electron scattering properties of the dosemeter and the phantom material. Some suggestions are presented for a different cavity theory for flat dosemeters dealing also with these phenomena. It describes the results to about the same degree of approximation as the Burlin theory, and fails to explain the observed energy dependence for electrons.